WEIGHT VISUALIZER for load cells
Indicator of weight with input for cargo cells or input 0-10V or 4-20mA.
Alphanumeric display to 5 characters / digits, height 100 mm, brightness automatically adjusted and suitable for external installation.
Container in stainless steel and watertight logons.
Configuration and calibration from remote control, serial trasmition Rs232, Rs485. RF868Mhz extern interface (optional).
Download installation and usage manual
Click to enlarge image
Power supply | 100 - 240 V ac – 50 / 60 Hz | |
Max power consumption | 25 W | |
Operating temperature | -10 / +40°C (14 °F / 104°F) Humidity maximum 85% non-condensing | |
Storage temperature | -20 / +50°C (-4 / +122°F) | |
Display | Alphanumeric 5-digit dot matrix 5 x 7 high-efficiency red LEDs with a wide viewing angle | |
Reading distance | Till at 35 m | |
Display brightness | Anti-reflection filter and automatic brightness adjustment in ambient light function. Suitable for direct light installation | |
Overall dimensions | 500 x 185 x 115 mm / 19.7 x 7.3 x 4.5 in (l x h x p) | |
Mounting | Wall-mounted with 4 external side holes Ø 6,5 mm with center distance 510 x 155 mm (l x h) | |
Case material | Stainless steel protected against dust | |
Protection level | IP 65 | |
Connections | Waterproof circular connectors with conductor wiring with screw terminal | |
Power supply load cells | 5 V dc / 120 mA (maximum 8 cells with 350 Ω in parallel) protected against short-circuit | |
Linearity | < 0.01% of full scale | |
Temperature deviation | <0,0003% of full scale/°C | |
Internal resolution | 24 bit | |
Weight resolution | Up to 60,000 divisions on the load capacity | |
Measuring range | -1.5 mV/V - +3.5 mV/V | |
Weight acquisition frequency | 6 Hz - 25 Hz | |
Digital filter | 0.2 Hz - 25 Hz | |
Number of weight decimals | from 0 to 3 decimal digit | |
Communication ports | RS232c / RS485 standard; 4-key remote control for configuration and operational commands | |
Max. cable length | 15m (Rs232c) and 1000m (Rs485) | |
Serial protocols | ASCII + programmable custom protocols | |
Baud rate | From 1200 to 115200 selectable | |
Remote control | Remote control with 4 key for configuration and operating commands | |
Options | External 868 MHz RF interface, Ethernet protocol, Profibus protocol | |
Conforms to standards | EN61000-6-2, EN61000-6-3, EN61010-1 |
Type | J1 5 poles serial female |
Length | ND |
1 | GND + Shield |
2 | RS232 Rx |
3 | RS485 - |
4 | RS485 + |
5 | RS232 Tx |
Type | J2 5 poles male load cell |
Length | ND |
1 | + Exc. Load cell |
2 | + signal load cell |
3 | - Signal load cell |
4 | - Exc. Load cell |
5 | Shield |
General precautions to be followed in the installation and assembly of load cells: